| Model Number |
VTICMO13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
05/24/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturer narrative - secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #(b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.In a separate visit the lens was exchanged for a replacement lens of different power.The problem resolved.Cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: g4- premarket identification: device bla: p030016 to pma/510(k): p030016, initially reported pma/510(k) number in the wrong field.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6 device evaluation: lens was returned dry in microcentrifuge vial.Visual inspection found the haptics deformed/stretched out.Dimensional and functional inspection found the lens to be within specifications.(b)(4).
|
| |
|