MAUDE Adverse Event Report: SYNTHES GMBH 2.7MM TI CORTEX SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 20MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.7MM TI CORTEX SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 20MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 402.880TS
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2024, the patient underwent an osteosynthesis with a plate for a fracture of left clavicle diaphysis.The surgeon inserted the screw in question to the variable angle (va) locking hole for reduction.The screw head broke off due to over-tightening.The shaft part of the screw was left in the body at the surgeon¿s discretion.Fixation was completed successfully.The surgery was completed successfully with no surgical delay.Patient was stable; there was no patient consequence.This report is for a 2.7mm ti cortex screw slf-tpng with t8 stardrive recess 20mm.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a synthes employee.H3, h6: a manufacturing record evaluation was performed for the finished device.Product code: 402.880ts.Lot: 5681p29.It was electronically reviewed and no nonconformance's / manufacturing irregularities were identified during the manufacturing process.The product was released on: 12-may-2023, manufacturing site: jabil grenchen, expiry date: 01-may-2028.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.H6 component code: g07002 ¿ device not returned.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM TI CORTEX SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 20MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK GRENCHEN (CH)
solothurnstrasse 186
grenchen 2540
SZ   2540
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19750750
Report Number8030965-2024-08770
Device Sequence Number17765137
Product Code HWC
UDI-Device Identifier10886982206749
UDI-Public(01)10886982206749
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112583
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number402.880TS
Device Lot Number5681P29
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/20/2024
Initial Report FDA Received Date07/16/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VA-LCP CLAVIC PL 2.7 SHAFT CS3 LE TAN
Patient SexUnknown
-
-