An error message was reported with the abbott diabetes care (adc) application in use with iphone (unspecified model), with ios 17.5.1, and with a software version 2.7.5.4061.The customer received a "replace sensor" error message and was unable to obtain readings.As a result, the customer experienced symptoms described as "difficulty eating and seeing", and a loss of consciousness.The customer was unable to self-treat, requiring non-healthcare professional (non-hcp) administration of "7-up and banana" for treatment.There was no report of death or permanent impairment associated with this event.
|
|
Software: the most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The customer experienced a "replace sensor" error message with the freestyle libre 2 application.Smartphone compatibility with the use of freestyle librelink and the iphone (unknown model), ios 17.5.1 has not been tested at the time of the investigation.The compatibility guide states that abbott diabetes care has not assessed compatibility on phone/operating systems other than those listed.This guide is available to the user on the websites where the product is launched.The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.Sensor: the most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|