• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. HYDROSOFT-3D-AV; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. HYDROSOFT-3D-AV; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 7110-0204-3D-A2
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Aneurysm (1708); Intracranial Hemorrhage (1891)
Date of Event 02/23/2021
Type of Reportable Event Serious Injury
Event or Problem Description
As reported through the rage clinical study: four days post-op, patient experienced increased lethargy, prompting ct scan which showed increase in hemorrhage in right frontal lobe.A repeat ct several hours later was stable.Patient was intubated for airway protection.Neurological exam continued to decline and patient was taken for angiogram, where he underwent balloon angioplasty and placement of a pipeline stent.No residual aneurysm was found during procedure.Recovery was complicated by vasospasm.Patient's exam improved after vasospasm treatments and he was extubated approximately 1.5 weeks after patient experienced increased lethargy.Patient returned to post-coiling neurologic baseline two days later.Adverse event was resolved without sequelae.An independent reviewer (cea) determined that the reported event was probably related to the study device and definitely related to study procedure.
 
Additional Manufacturer Narrative
Items returned: n/a.Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging was not provided for this investigation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform an analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review: potential complications: potential complications include but are not limited to: hematoma at the site of entry, vessel perforation, aneurysm rupture, parent artery occlusion, incomplete aneurysm filling, emboli, hemorrhage, ischemia, vasospasm, coil migration or misplacement, premature or difficult coil detachment, clot formation, revascularization, post-embolization syndrome, and neurological deficits including stroke and possibly death.Cases of chemical aseptic meningitis, edema, hydrocephalus and/or headaches have been associated with the use of embolization coils in the treatment of large and giant aneurysms.The physician should be aware of these complications and instruct patients when indicated.Appropriate patient management should be considered.Warnings and precautions the hes is intended for single use only.Do not resterilize and/or reuse the device.After use, dispose in accordance with hospital, administrative and/or local government policy.Do not use if the packaging is breached or damaged.If a coil must be retrieved from the vasculature after detachment, do not attempt to withdraw the coil with a retrieval device, such as a snare, into the delivery catheter.This could damage the coil and result in device separation.Remove the coil, microcatheter, and any retrieval device from the vasculature simultaneously.Detachment of the hes coil.33.When the v grip® detachment controller is properly connected to the v trak® delivery pusher, a single audible tone will sound and the light will turn green to signal that it is ready to detach the coil.If the detachment button is not pushed within 30 seconds, the solid green light will slowly flash green.Both flashing green and solid green lights indicate that the device is ready to detach.If the green light does not appear, check to ensure that the connection has been made.If the connection is correct and no green light appears, replace the v grip® detachment controller.35.Push the detachment button.When the button is pushed, an audible tone will sound and the light will flash green.36.At the end of the detachment cycle, three audible tones will sound and the light will flash yellow three times.This indicates that the detachment cycle is complete.If the coil does not detach during the detachment cycle, leave the v grip® detachment controller attached to the v trak® delivery pusher and attempt another detachment cycle when the light turns green.37.The light will turn red after the number of detachment cycles specified on the v grip® labeling.Do not use the v grip® detachment controller if the light is red.Discard the v grip® detachment controller and replace it with a new one when the light is red.38.Verify detachment of the coil by first loosening the rhv valve, then pulling back slowly on the delivery system and verifying that there is no coil movement.If the implant did not detach, do not attempt to detach it more than two additional times.If it does not detach after the third attempt, remove the delivery system.The physical device was not available for evaluation to determine if a condition existed that would have caused or contributed to event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.This information may be updated if additional information is provided at a later date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROSOFT-3D-AV
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key19753162
Report Number2032493-2024-00542
Device Sequence Number9004891
Product Code KRD
UDI-Device Identifier00816777025388
UDI-Public(01)00816777025388(11)171025(17)220930(10)171025525
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model Number7110-0204-3D-A2
Device Lot Number171025525
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2024
Initial Report FDA Received Date07/16/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/25/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age47 YR
Patient SexMale
-
-