| Model Number |
VICM5_13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Toxic Anterior Segment Syndrome (TASS) (4469)
|
| Date of Event |
06/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon performed bilateral sequential surgery and implanted a 13.2 vicm5; -5.0 d implantable collamer lens (icl) into the patient`s left eye (os) on (b)(6) 2024.Concomitant products used intraoperatively included opegan viscoelastic and the msi injector system.Toxic anterior segment syndrome (tass) was diagnosed based on clinical signs.Additional treatment was prescribed.Per report from 10-june-2024, patient condition was bcva:20/13; ucva:20/20; iop:14mmhg.Inflammation was graded as not severe but patient was under careful observation.According to the report the cause of the event was unknown.Lens remains implanted.Additional information has been requested but none has been forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: health effect - impact code (f): additional medications: 4644 - psl, betamethasone valerate/ gentamicin sulfate ophthalmic drops, tropicamide phenylephrine hydrochloride ophthalmic drops (mydrin-p ophthalmic solution).H6 - work order search: no similar complaint was reported for units within the same lot.H11: manufacturer narrative: a review of the device labeling was completed.Intraocular inflammation is identified in the labeling as known potential adverse event following icl implantation.The dfu provides the surgeon instruction for complete ovd removal.Precaution: (6) after inserting this product, aspirate the viscoelastic substance completely from inside the eye.Do not use highly viscous viscoelastic substances that are difficult to aspirate completely.Toxic anterior segment syndrome (tass) is an acute sterile postoperative inflammation that can occur after uncomplicated or complicated intraocular surgery.While improper sterilization and contamination of surgical instruments remains the most common risk factor associated with tass, ocular viscoelastic, and improper preparation of antibiotics for intracameral injection may also lead to tass.An exact cause could not be determined as the event could be multifactorial in nature including patient and/or procedure related factors.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h6: device history record (dhr) review: the dhr review indicated that the product has been manufactured within the established process parameters and that there is no indication that the manufacturing and/or processing of the device contributed to the complaint issue.Corrected data: d4: primary unique device identifier (udi) #: (b)(4)."udi related data quality updates only" h4: device manufacture date: 07-nov-2023 claim # (b)(4).
|
| |
|
Search Alerts/Recalls
|