| Model Number |
720182-01 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this inflatable penile prosthesis (ipp) was not performing as expected.Two weeks later a revision surgery was performed, upon examination a hole in the reservoir was found.The pump was removed and replaced, while the existing reservoir was kept in the patient and a new one was added to the system.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this inflatable penile prosthesis (ipp) was not performing as expected.Two weeks later a revision surgery was performed, upon examination a hole in the reservoir was found.The pump was removed and replaced, while the existing reservoir was kept in the patient and a new one was added to the system.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly analyzed.The momentary squeezed pump was microscopically inspected, and the tubing was cut down to the pump block, the tubing was not returned for analysis.The pump passed the leakage and functional testing.The investigation conclusion code of cause not established was chosen.
|
| |
|