• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 720182-01
Medical Device Problem Codes Mechanical Problem (1384); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/12/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this inflatable penile prosthesis (ipp) was not performing as expected.Two weeks later a revision surgery was performed, upon examination a hole in the reservoir was found.The pump was removed and replaced, while the existing reservoir was kept in the patient and a new one was added to the system.No patient complications were reported.
 
Event or Problem Description
It was reported that this inflatable penile prosthesis (ipp) was not performing as expected.Two weeks later a revision surgery was performed, upon examination a hole in the reservoir was found.The pump was removed and replaced, while the existing reservoir was kept in the patient and a new one was added to the system.No patient complications were reported.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly analyzed.The momentary squeezed pump was microscopically inspected, and the tubing was cut down to the pump block, the tubing was not returned for analysis.The pump passed the leakage and functional testing.The investigation conclusion code of cause not established was chosen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS
Common Device Name
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key19757627
Report Number2124215-2024-44153
Device Sequence Number17765205
Product Code FHW
UDI-Device Identifier00878953005676
UDI-Public00878953005676
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/21/2022
Device Model Number720182-01
Device Catalogue Number720182-01
Device Lot Number0178273004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2024
Supplement Date Received by Manufacturer09/12/2024
Initial Report FDA Received Date07/16/2024
Supplement Report FDA Received Date10/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CX 18CM: UPN: 72404273 / LOT: 0179005007; CX 18CM: UPN: 72404273 / LOT: 0179005007; PUMP: UPN: 72404310/ LOT: 0178455001; PUMP: UPN: 72404310/ LOT: 0178455001; RTE: UPN: 72404321 / LOT: 0134416015; RTE: UPN: 72404321 / LOT: 0134416015
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age71 YR
Patient SexMale
-
-