• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTERA ONCOLOGY, INC INTERA REFILL KIT; HUBER NEEDLE WITH INFUSION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTERA ONCOLOGY, INC INTERA REFILL KIT; HUBER NEEDLE WITH INFUSION SET Back to Search Results
Model Number AP07014US
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by a healthcare provider that they had 'an ongoing issue with leaking between the intera needle and connector set.' healthcare provider stated 'the connections being very thoroughly tightened, leaking persists when medication is instilled.Healthcare provider asked if the female end [of the needle] is not circular but straight on either side if this could be resulting in a connection that's not secure.
 
Additional Manufacturer Narrative
The suspect device was not made available for return and evaluation.Other devices from this lot have been previously evaluated.Three devices from the same lot were pulled from inventory and evaluated by intera engineering.In summary, the engineer pressurized the connection between the needle and tubing set and exerted maximum manual force.No signs of leakage were observed.This was repeated for a total of 3 times on 3 different kits from the same lot.Additional 3rd party testing for leakage was conducted, and acceptable device performance (without leakage) was confirmed.Therefore, the origin of the complaint is unconfirmed.Blank fields in the mdr form represent unknown information.If additional information is received at a later date, a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERA REFILL KIT
Common Device Name
HUBER NEEDLE WITH INFUSION SET
Manufacturer (Section D)
INTERA ONCOLOGY, INC
180 wells avenue
ste 300a
newton MA 02459
Manufacturer (Section G)
INTERA ONCOLOGY, INC
180 wells avenue
ste 300a
newton MA 02459
Manufacturer Contact
sarah lapp
180 wells avenue
ste 300a
newton, MA 02459
7814895724
MDR Report Key19757764
Report Number3015537318-2024-00060
Device Sequence Number15704543
Product Code PTI
UDI-Device Identifier00850014110079
UDI-Public00850014110079
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K213823
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/09/2026
Device Model NumberAP07014US
Device Catalogue NumberAP-07014US
Device Lot Number23J038CT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2024
Initial Report FDA Received Date07/16/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/09/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-