This report is being supplemented to provide additional information based on the approved final investigation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the results of the investigation, it was presumed that the suggested event occurred due to breakage of the image sensor unit.However, the definitive root cause of the reported event could not be determined, and the device did not meet the specifications, thereby confirming the occurrence of the event.The event can be detected/prevented by following the instructions for use in: precaution for disappeared or frozen endoscopic image; warning: follow the precautions given below when handling the instrument.Otherwise, the endoscopic image may disappear unexpectedly, or the frozen image may not be restored during the examination.Precaution for disappeared or frozen endoscopic image; caution: turn the video system center off before connecting or disconnecting the videoscope cable from the electrical connector on the endoscope.Turn the switch on or off only when the videoscope cable is connected to both the video system center and electrical connector on the endoscope.Failure to do so can result in equipment damage, including destruction of the ccd.Do not touch the electrical contacts inside the electrical connector.Ccd damage may result.3.6 inspection of the endoscopic system inspection of the endoscopic image.4.While observing the palm of your hand, confirm that the wli and nbi endoscopic image is free from noise, blur, fog, or other irregularities.5.Angulate the bending section and confirm that the endoscopic image is free from momentary disappearing or other irregularities.Chapter 5 troubleshooting: if the endoscope is visibly damaged, does not function as expected or is found to have irregularities during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope.Contact olympus.Some problems that appear to be malfunctions may be correctable by referring to section 5.1, ¿troubleshooting guide¿.If the problem cannot be resolved by the described remedial action, stop using the endoscope and send it to olympus for repair.Olympus does not repair accessory parts.If an accessory part becomes damaged, contact olympus to purchase a replacement.Olympus will continue to monitor field performance for this device.
|