| Catalog Number |
23533903 |
| Medical Device Problem Codes |
Degraded (1153); Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported: "the nail adapter 2353-3903 must be damaged.It would not allow the nail to be threaded on to the assembly with the holding screw 2353-3904.There was no replacement so the surgeon implanted the nail and had to ¿free hand¿ the distal screws.We tried both nail holding screws but they weren't able to start threading in to the nail due to the nail adapter.There is no visible damage to the holding screw threads or any other part of the screw.There may be damage to the nail adapter where it connect to the nail but i don't have a second one to compare it to.The surgeon ended up bypassing the nail adapter and targeting jig and used the "freehand" technique to drill and insert the two distal screws.He was satisfied with the results.".
|
| |
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed, since the returned device was found to be functional and conforming to specifications.The device inspection revealed the following: the received nail adapter shows typical signs of normal usage.No significant damage/deformation observed other than the pegs which were slightly rounded at the edges which is normal.Overall, the device looked in a good condition.The returned nail adapter and nail holding screws were assembled with a sample nail.The assembly could be done easily with no obstruction in inserting the nail holding screw in the adapter and then attaching the nail.This could be confirmed with both of the nail holding screws.Thus, the reported issue could not be reproduced and hence could not be confirmed.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on available information, the root cause of issue is deemed to be user related.The reported issue could not be reproduced.The devices were working as intended.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "the nail adapter 2353-3903 must be damaged.It would not allow the nail to be threaded on to the assembly with the holding screw 2353-3904.There was no replacement so the surgeon implanted the nail and had to ¿free hand¿ the distal screws.We tried both nail holding screws but they weren't able to start threading in to the nail due to the nail adapter.There is no visible damage to the holding screw threads or any other part of the screw.There may be damage to the nail adapter where it connect to the nail but i don't have a second one to compare it to.The surgeon ended up bypassing the nail adapter and targeting jig and used the "freehand" technique to drill and insert the two distal screws.He was satisfied with the results.".
|
| |
|