• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD DOUBLE BEND, 75 CM; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD DOUBLE BEND, 75 CM; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 1457Q/75
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 06/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had pocket infection and erosion.The implant site was red.Upon further investigation via trans-esophageal echocardiogram, possible vegetation was noted on an unidentified lead.The entire pacemaker system was explanted.The patient was in stable condition.
 
Additional Manufacturer Narrative
A device history record (dhr) review was performed, and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The cause of infection could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD DOUBLE BEND, 75 CM
Common Device Name
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19761065
Report Number2017865-2024-60421
Device Sequence Number20984895
Product Code OJX
UDI-Device Identifier05414734510172
UDI-Public(01)05414734510172(10)A000091683(17)221231
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number1457Q/75
Device Lot NumberA000091683
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2024
Supplement Date Received by Manufacturer08/28/2024
Initial Report FDA Received Date07/17/2024
Supplement Report FDA Received Date08/28/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexFemale
Patient Weight64 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-