MAUDE Adverse Event Report: BECTON DICKINSON TRAY SPN WHIT25G3.5 L/B-D/E PLAST DRAPE; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON TRAY SPN WHIT25G3.5 L/B-D/E PLAST DRAPE; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 405671
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event or Problem Description
Material#:405671.Batch#: 0001544106.It was reported by the customer that the medication in the spinal tray was failing.Crna brought us 3 trays with the same lot numbers that anesthesia had noted to have failed during several previous cases this pr is created based on the follow-up response from the customer on 03-jul-2024.Event details: clinicians are reporting 6 total incidents now: 3 events with undetermined dates, and a single event on each of these dates: 5-29-24; 6-20-24; 6-26-24.On 15jul2024: dchu followed up directly with the customer via telephone.It was confirmed the lot involved in these event dates is the same as previously reported, lot 0001544106.At this time the facility cannot clearly identify the results of each of these events, there has been no patient impact as far as the reporter was aware but he states that it cannot be confirmed if these procedures were completed using a new vial or if they had to revert to general anesthesia.Products will be returned.No additional information to report at this time.Polc.
 
Additional Manufacturer Narrative
(b)(4) follow up mdr for device evaluation: ten trays were provided to our quality team for investigation.The product was visually inspected, no issues were identified within any of the ampules.A review of the internal manufacturing device records and raw material history files for the reported lot number 0001544106 was performed and no recorded quality problems or rejections related to this incident were found, all procedural and functional requirements for product release have been met.The sterilization record was reviewed, no issues were identified.Based on the available information we are not able to identify a root cause related to our process at this time.The medication within the kit is provided by a supplier.We have notified the supplier of this incident.Complaints received for this device and reported condition will continue to be monitored by our quality team for signs of emerging trends.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAY SPN WHIT25G3.5 L/B-D/E PLAST DRAPE
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key19761912
Report Number1625685-2024-00078
Device Sequence Number20960581
Product Code CAZ
UDI-Device Identifier00382904056711
UDI-Public(01)00382904056711
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Catalogue Number405671
Device Lot Number0001544106
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2024
Supplement Date Received by Manufacturer09/30/2024
Initial Report FDA Received Date07/17/2024
Supplement Report FDA Received Date10/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-