MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE Back to Search Results
Model Number DBEC-125
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Myocardial Infarction (1969)
Date of Event 12/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6 investigation conclusion code 22: the diamondback 360® coronary orbital atherectomy system instructions for user manual states that a myocardial infarction is a potential adverse event that may occur and/or require intervention with use of the system.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Event or Problem Description
The target treatment area was a 4mm, 75% stenosed, heavily calcified mid right coronary artery (rca).The diamondback 360 coronary orbital atherectomy device (oad) was spun for a total of 11 times for treatment, 8 times on low speed and 3 times oh high speed.Angioplasty and stent placement were performed to complete the procedure.The clinical event committee reviewed procedural images and concluded that the diamondback coronary orbital atherectomy device possibly contributed to a myocardial infarction.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM
Common Device Name
CORONARY ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
poonoy chanthavongsa
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key19762114
Report Number3004742232-2024-00239
Device Sequence Number8604854
Product Code MCX
UDI-Device Identifier10850026568087
UDI-Public(01)10850026568087(17)240930(10)450674-1
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P130005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberDBEC-125
Device Catalogue Number7-10060-09
Device Lot Number450674-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2024
Initial Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight100 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-