MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507490MC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
05/18/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: silent rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female who underwent a breast augmentation primary with a mentor memorygel, 510cc on the left, and 490cc on the right side, silicone prosthesis experienced bilateral silent rupture post operation.The patient went into surgery for a bigger set of implants and during surgery the md/surgeon discovered bilateral rupture.As a result, patient underwent bilateral replacements as follow: left catalog number 3505700bc, serial number (b)(6) right catalog number 3505700bc, serial number (b)(6) on june 22, 2024.This report relates to the right side.
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on august 12, 2024: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device visual analysis of the returned sample revealed that the sm mod classic gel, 490cc breast implant was found to have a tear measuring 1.0 cm.On the posterior view.Additionally, an area of shell abrasion was observed on the edges of the tear.The evaluation determined that the possible cause of the rupture was consistent with normal wear.Shell abrasion suggests in-vivo folding or creasing of the device.This may be the result of the following factors: continuous and sustained stresses to the device such as too small breast pocket and folding or wrinkling of the shell in the breast pocket.In some cases, the breast implants may also wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Manufacturer¿s reference number: (b)(4).
|
| |
|