|
H10: h3, h6: the reported device was received for evaluation.A visual inspection revealed that it is not in its original packaging.The threads near the proximal end of the device were fractured away.There is biological debris on the returned device.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor material specification found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Based on this investigation, the need for corrective action is not indicated.
|