No device was returned for evaluation.Additionally, no photographs were provided for review.Operative notes were not provided for review of usage/technique.Based on the information obtained, the root cause of the reported event is unable to be determined conclusively; however, off-axis excessive force and/or excessive twisting motions placed on the device during use or wear/fatigue of the device during the over 10 years of field use may have been causes or contributing factors.Labeling review: ".Residual risks to the patient associated with use of general surgical instruments are: instrument malfunctions which potentially result in surgical delays (thereby causing additional exposure to anesthesia, blood loss, and infection), a change in surgical plan, failure to achieve optimal clinical result and/or, infrequently, cancellation of a procedure.Such malfunctions include instrument fracture, which could make necessary removal difficult or sometimes impossible, with possible consequences of late infection and migration; instrument fracture may also result in injury to the patient.".".Residual risks to surgical staff associated with use of general surgical systems are: instrument failures or malfunctions which prevent intended use, resulting in decreased operational efficiency and general inconvenience.".".Pre-operative warnings: the method of use for the instruments are to be determined by the user¿s experience and training in surgical procedures.The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury.".".Intra-operative warnings: the physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted." if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|