• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SALTER LABS INTUBRITE; MAC 4 SINGLE USE DISPSOABLE LARYNGOSCOPE BLADE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SALTER LABS INTUBRITE; MAC 4 SINGLE USE DISPSOABLE LARYNGOSCOPE BLADE Back to Search Results
Model Number 1024.C
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is reported to be available but has not yet been received by the manufacturer.A review of the device history record is in-progress.All information reasonably known as of 17 jul 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
 
Event or Problem Description
Sunmed holdings llc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the first of two reports.Refer to 3000219639-2024-00064 for the second report.It was reported, the physician attempted to intubate the patient; however, the light on the blade did not work; the blade was replaced with one that worked.There was no reported injury to the patient.It was additionally reported, the defective blade was tested after the patient was stable; it worked several times but also shorted out, as it would "come on and go back off.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTUBRITE
Common Device Name
MAC 4 SINGLE USE DISPSOABLE LARYNGOSCOPE BLADE
Manufacturer (Section D)
SALTER LABS
30 spur dr
el paso TX 79906
Manufacturer (Section G)
SALTER LABS
30 spur dr
el paso TX 79906
Manufacturer Contact
racheal cooper
2710 north ridge dr nw
ste a
grand rapids, MI 49544
2054826767
MDR Report Key19767586
Report Number3000219639-2024-00063
Device Sequence Number12912773
Product Code CCW
UDI-Device Identifier00607411964664
UDI-Public(01)00607411964664
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1024.C
Device Lot Number2022-01-01-CMC4
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2024
Initial Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-