|
Additional information is provided in sections d.4, d.9, h.3, h.4, h.6 and h.11.One opened i/a tip was returned for the reported issue of occlusion (blocked).The returned sample was visually inspected and was found to be conforming.The sample was then functionally tested for occlusion/leakage and was found to be conforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned sample was found to be visually and functionally conforming; therefore, a blocked tip as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the tip was manufactured to specifications.All i/a tips are 100% visually inspected prior to being released from the manufacturing facility.The manufacturer internal reference number is: (b)(4).
|