• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THOMMEN MEDICAL AG ELEMENT IMPL., PF 5.0, E Ø 5.0, L 9.5; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THOMMEN MEDICAL AG ELEMENT IMPL., PF 5.0, E Ø 5.0, L 9.5; DENTAL IMPLANT Back to Search Results
Model Number 4.13.942
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 07/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Implant didn't achieve osseointegration, primary stability was achieved, implant was completely covered with bone, no thread tap used, smoker, infection, abscess.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEMENT IMPL., PF 5.0, E Ø 5.0, L 9.5
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH  2540
Manufacturer (Section G)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH   2540
Manufacturer Contact
astrid weber
neckarsulmstrasse 28
grenchen 2540
CH   2540
MDR Report Key19768630
Report Number3003184527-2024-02069
Device Sequence Number18287364
Product Code DZE
UDI-Device Identifier07640156471366
UDI-Public07640156471366
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072649
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 07/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number4.13.942
Device Lot Number25539
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2024
Initial Report FDA Received Date07/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age51 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-