Model Number CNWTT0
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for analysis.The root cause for the reported complaint could not be determined as no sample was returned for analysis.The product was subject to handling and the reported damage was highlighted post implantation.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A healthcare professional reported that after an intraocular lens (iol) implantation procedure, scratches on the posterior side of the lens was observed.There was no patient harm.Additional information was requested and received stating there was no issue with the iol as well as the delivery system.No further reports required.Additional information was received stating the operation was completed, there was no reason for explantation - minimum optic damage comparable to a defect after ndyag, moreover far enough from the optical axis not to interfere.There are two medical device reports associated with this patient.This report is associated with the right eye.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TRIFOCAL HYDROPHOBIC INTRAOCULAR LENS
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key19770380
Report Number9612169-2024-00714
Device Sequence Number18372862
Product Code MFK
UDI-Device Identifier00380652408042
UDI-Public00380652408042
Combination Product (Y/N)N
Initial Reporter CountryEZ
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/07/2026
Device Model NumberCNWTT0
Device Catalogue NumberCNWTT0.160
Device Lot Number25424230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/02/2024
Initial Report FDA Received Date07/18/2024
Date Device Manufactured06/09/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE; VISCOAT OPHTHALMIC VISCOSURGICAL DEVICE
Patient SexUnknown
-
-