• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MEDICAL PRODUCTS LIGASURE ATLAS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MEDICAL PRODUCTS LIGASURE ATLAS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LS1037
Medical Device Problem Codes Break (1069); Peeled/Delaminated (1454); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, prior to use, two units of ls1037 got the same issue; insulation got removed after inserting the instrument in the generator's port.On the photo provided, it can also be seen that the wire on the jaw was exposed and broken.There was no patient involvement.Medtronic's initial evaluation of the incident device found the ligasure devices have insulation damage.There was an evidence of use and the portion of the insulation of the wire on the ligasure devices were not returned.
 
Additional Manufacturer Narrative
Concomitant medical productst: ls1037, ls1037 ligasure atlas handswitch37cm, (lot #s22k0004).H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the jaw wire's insulation was damaged.The jaw wire for this device was severed.It was reported that the insulation was damaged, wire was severed, and wire was exposed.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur during the insertion/extraction from a trocar.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the jaw wire's insulation was damaged.A portion of the device's insulation was not returned.It was reported that the insulation was damaged, wire was severed, and wire was exposed.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: carefully insert and withdraw instruments from cannulas (trocars) to avoid possible damage to the devices and/or injury to the patient.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a procedure, two units of ls1037 got the same issue, insulation got removed after inserting the instrument while inserting into the generator's port.None of the piece fell into the patient's cavity.On the photo provided it can also be seen that the wire on the jaw was exposed and broken.There was no patient injury.Medtronic's initial evaluation of the incident device found the ligasure devices have insulation damage.There was an evidence of use and the portion of the insulation of the wire on the ligasure devices were not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE ATLAS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CH  201114
Manufacturer (Section G)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CH   201114
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19771526
Report Number3006451981-2024-00097
Device Sequence Number12800513
Product Code GEI
UDI-Device Identifier10884524001630
UDI-Public10884524001630
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K010013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLS1037
Device Catalogue NumberLS1037
Device Lot NumberS22K0004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2024
Supplement Date Received by Manufacturer10/30/2024
Initial Report FDA Received Date07/18/2024
Supplement Report FDA Received Date11/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.