This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: investigation summary = >the actual device was returned for evaluation.During repair, an evaluation was performed; and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component wear.
|
|
It was reported from india that during service and evaluation, it was determined that the quick coupling for k-wires device had the following failures: frozen/will not move and cannot engage attachment - coupling.It was further determined that the device failed pretest on the following: check function of adjusting sleeve and check the tool quick coupling.It was noted in the service order that the device was not working.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown but was noted to have occurred in 2024.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
|