| Catalog Number |
544965 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "the jaws got misaligned during a surgery.Therefore, the applier was replaced with another one to complete the procedure.No clip fell/remained in the patient and no injury to the patient occurred".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Other remarks: n/a, corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that "the jaws got misaligned during a surgery.Therefore, the applier was replaced with another one to complete the procedure.No clip fell/remained in the patient and no injury to the patient occurred".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The sample was returned to the manufacturer for evaluation.The manufacturer reported: "the dhr for the returned instrument was reviewed and found complete without any irregularities.This instrument was produced at the tecomet, kenosha, wi facility as part of a (b)(4) pc.Lot in october of 2022.It was found that this order was made from the correct materials and components and all approved processes were followed.Evaluation of the returned instrument as received shows that as received it is missing its luer flush port cap and the handle to jaw and knob rotation mechanisms are dry and sluggish feeling and in need of proper lubrication.Further evaluation shows that the jaws are slightly loose and misaligned in the closed position.We are able to validate this complaint for the returned instrument.After the initial evaluation this instrument was dis-assembled in order to evaluate its internal components and it was found that the inner drive rod (n00185) bosses are both damaged where they engage the jaws.We suspect that the damaged drive rod bosses caused the jaws to become slightly loose and misaligned in the closed position.We are unable to determine what caused the drive rod bosses to become damaged but mishandling such as excessive force being applied to the handle to jaw mechanisms of this device at the end user's facility is suspected.All (b)(4) instruments from this lot were 100% visually inspected and function tested and properly lubricated prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed." teleflex will continue to monitor and trend on complaints of this nature.
|
| |
|