• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-D; Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-D; Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation Back to Search Results
Model Number A-TCSE-DD
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One bi-directional, curve d-d, tacticath sensor enabled contact force ablation catheter was received for evaluation.Electrodes 1 and 2 read as open circuits during electrical testing, consistent with the reported event.Further investigation revealed that the catheter shaft material had been fractured between electrode rings 2 and 3.Conductor wires 1 and 2 were determined to be fractured at the location of the fracture in the shaft material, consistent with the detected open circuits, and with the reported event.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
 
Event or Problem Description
This report is to advise of a fracture found in the catheter shaft during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-D
Common Device Name
Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
wendy mccarty
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key19776055
Report Number3008452825-2024-00385
Device Sequence Number18225889
Product Code OAE
UDI-Device Identifier05415067027634
UDI-Public(01)05415067027634(10)10093594(17)250430
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P130026
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberA-TCSE-DD
Device Catalogue NumberA-TCSE-DD
Device Lot Number10093594
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2024
Initial Report FDA Received Date07/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-