EXACTECH, INC. TRULIANT TIB IMP CRC INSERT SZ 4, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-022-51-4011 |
| Medical Device Problem Codes |
Unstable (1667); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: concomitant devices: (b)(6) 02-020-13-0240 - truliant cr cem fem cr cem left sz 4, (b)(6) 200-07-35 - advanced patella 35mm 3 peg implant, (b)(6) 02-022-45-4040 - truliant tib fit tray cem sz 4f / 4t.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 44 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, permanent and debilitating injuries and damages, significant pain and discomfort, gait impairment, poor balance, difficulty walking; component part loosening, soft tissue damage, bone loss.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Multiple mdr reports were filed for this event, please see associated report: 1038671-2024-02450.This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: h6 (type of investigation).The following sections were corrected: h6 (health effect - clinical code, medical device problem code).The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|