| Model Number |
CR2 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device was missing the magnet from the lid of the device.In this state the device will not turn on automatically when the lid is activated, which can cause a delay in defibrillation therapy if needed or may lead to use error resulting in a failure to deliver defibrillation.There was no patient use reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer received a replacement lid and battery.The device was not returned to stryker for evaluation.The root cause of the reported issue could not be determined.
|
| |
|
Search Alerts/Recalls
|