|
A facility reported a microsensor (id 826653) was implanted on (b)(6) 2024.On (b)(6) 2024, a minimal amount of cerebrospinal fluid (csf) leakage was discovered in the connector.Therefore, the physician removed the sensor and stopped monitoring.The sensor was used with cp express monitor (id 826635) and icp express cable (id 826636).
|
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The microsensor (id 826653) was returned for evaluation.Device history record (dhr) -the batch record was reviewed for any anomalies or non-conformances (ncs) related to the finished device, and none were identified, related to this report.Failure analysis - the device was evaluated for leaking.No issues were identified with the device.The complaint could not be confirmed.Root cause analysis - the potential root cause includes the use error and the lure lock not sealing.
|