• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION ADAPTER C100; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION ADAPTER C100; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM Back to Search Results
Catalog Number 515306
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
Another incident has occurred.Visible in attached photos other pirs: (b)(4).(b)(6) are experiencing a "foreign matter" in the iv bags, it looks small and white with a potential plastic nature.She is unsure where this is coming from, possibly phaseal? it's mainly with trastuzumab (grey vial bung) and rituximab (grey vial bung) but can occur with other drugs.Its happening in baxter pvc and non pivc bag.No drawing up needle used.They have a strict inspection policy for looking for foreign matter, in all cases it has not occured until they attach injector (515003) to the bag adapter (515306).So they check the syringe before injecting into bag and aren't seeing the matter.
 
Additional Manufacturer Narrative
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Additional Manufacturer Narrative
(b)(4) follow up mdr for device evaluation: multiple iv bags with the infusion adapter attached were provided to our quality team for investigation.Through visual inspection, particles were observed within the iv bag.The particles appeared white in color and based on visual characteristics, it is believed to be small particles of the phaseal membrane.Unfortunately, we were not able to safely extract the particles, therefore the specific composition cannot be identified at this time.Fragmentation testing is performed according to procedure, to evaluate any particulates generated by the injector when mated to other components after ten activations.As a lot number was unavailable for this incident, a device history record review could not be completed.While phaseal needles and spikes are designed to reduce coring, there are several factors that may impact coring tendency.Coring from the membrane may occur due to fragmentation caused by multiple injections or excessive welding of the membrane.Coring of the rubber stopper may result depending on the stopper quality, the design and dimensions of the needles used.Based on the available information, we are not able to identify a root cause related to the manufacturing process at this time.Complaints received for this device and reported condition will continue to be tracked and trended for future occurrence.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSION ADAPTER C100
Common Device Name
CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key19781955
Report Number3003152976-2024-00421
Device Sequence Number18159675
Product Code ONB
UDI-Device Identifier00382905153068
UDI-Public(01)00382905153068
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K123213
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number515306
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2024
Supplement Date Received by Manufacturer10/03/2024
Initial Report FDA Received Date07/19/2024
Supplement Report FDA Received Date10/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-