|
H3: an axium prime coil was returned for analysis.The headway 17 catheter used in the event was not returned.The axium prime coil was returned without introducer sheath.The axium prime push wire was found to be returned in two segments.The proximal segment was measured to be 6.6cm.The axium prime coil push wire was found to be broken distal to load indicator and bent at ~48.4cm from proximal end.The implant coil was found to be returned already detached.The shield coil was found to be intact.The implant coil was found to be stretched and damaged.All other subassemblies appeared to be normal, and no other anomalies were observed.The axium prime coil could not be used for resistance testing with an in-house micro catheter due to their damaged condition.Based on the device analysis and the reported information, the customers report of ¿coil resistance/stuck in cath¿ was confirmed.Possible causes for coil resistance in catheter are, but not limited to, use of an incompatible device for delivery, severely tortuous patient anatomy, failure to hydrate the axium coil prior to use, and lack of continuous flush during delivery.Based on the device analysis and the reported information, the customer¿s report of ¿pusher kink/damage¿ was confirmed as the axium pusher was found to be bent and the implant coil was found to be damaged; however, the cause of the damage/stretch could not be determined.Possible causes are coil is not retracted in a one-to-one motion with the implant pusher during repositioning, push wire rotation, user advances or retracts the coil against resistance.There were no reported issues pushing the axium implant coil from within the introducer sheath during preparation per ifu.It is possible the implant coil became damaged when retracting the implant coil following hydration or due to improper hubbing technique (over tightening of the rhv) subsequently causing the implant coil to become stuck.The headway 17 micro catheter has a labeled id (inner diameter) of 0.017¿, as per an online source.Per axium prime coil ifu (instructions for use): ¿axium prime detachable coils should be delivered through a microcatheter with a minimum inside diameter of 0.0165¿-0.017¿ with two marker bands.¿ therefore, the headway 17 micro catheter was found to be compatible for use with the axium prime coil and can be ruled out as a potential cause.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Medtronic received a report that the axium coil pusher distal segment was kinked.It was reported that the coil was kinked/damaged.There was friction or difficulty during testing or delivery.The continuous flush was administered during the procedure.The damage was located on the distal segment of the pushwire.The reported device and any accessory devices were prepared as indicated in the instructions for use (ifu).Ancillary devices include an excelsior sl-10 45 and headway 17 microcatheter, synchro 14 guidewire, and lvis evo.Additional information was received that vessel tortuosity was normal.The procedure was completed "good".
|