Investigation - evaluation: as reported, a 'universa soft ureteral stent set' was found creased and broke into two pieces prior to patient contact.The damage was found after cutting the tether off.The procedure was completed with a new stent.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), and quality control (qc) procedures were conducted during the investigation.Interview of personnel was also conducted.Functional tests and visual inspection of the returned complaint device was also conducted.A device failure analysis was able to be conducted as the device was returned for investigation.The device¿s coil had been separated from the stent body.Given the way the separation presented itself as well as the lack of attached tether, unintended user error during the tether removal process could not be ruled out as the root cause of the damage.Cook has concluded that the device was manufactured to specification.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no related non-conformances reported for lot.A complaint history database search showed no other related complaints associated with the failure mode for the complaint device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.Cook also reviewed product labeling; the product ifu, [t_usus_rev2] ¿universa soft ureteral stents and stent sets¿ contains the following information related to the reported failure mode: "how supplied: upon removal from the package, inspect the product to ensure no damage has occurred." based upon the available information and results of the investigation, cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|