OLYMPUS WINTER & IBE GMBH TELESCOPE "ULTRA", 5.4 MM, 45°; LAPAROSCOPE Back to Search Results
Model Number WA4KL545
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: (b)(6).The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported the light guide part of the rigid scope was missing.There were no reports of patient harm.
 
Event or Problem Description
The event occurred during postoperative device removal after therapeutic unspecified procedure that was completed with the same device.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over seven (7) years since the subject device was manufactured.The device was returned to olympus for inspection, and the customer reportable malfunction was confirmed.Based on the results of the investigation, the probable cause of the "light guide connector detached" malfunction would likely be related to user error, improper handling as well as application of excessive force.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE "ULTRA", 5.4 MM, 45°
Common Device Name
LAPAROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19789636
Report Number9610773-2024-01763
Device Sequence Number4145282
Product Code GCJ
UDI-Device Identifier04042761082891
UDI-Public04042761082891
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K150633
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA4KL545
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2024
Supplement Date Received by Manufacturer07/26/2024
Initial Report FDA Received Date07/22/2024
Supplement Report FDA Received Date08/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/06/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
WA03310A-LIGHT-GUIDE CABLE, 4.25 MM, 3 M, CF TYPE
Patient SexUnknown
-
-