| Catalog Number |
Y497G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.1.Any patient consequences? 2.Were there any unexpected outcomes or complications as a result of the prolonged surgery time? 3.Was there any change in the patient¿s post-operative care due to the prolonged procedure? a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a hip replacement on (b)(6) 2024 and suture was used.During the procedure, after closing the wound on the hip area, the suture broke before the knot could be tied.Another like device was used to close the wound.The surgery was delayed 10 minutes.The procedure was successfully completed.There were no adverse patient consequences.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 8/28/2024.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.One paper lid, one labeled winding former, and one suture piece were received for analysis.Product code y497g.During visual inspection of the returned sample, no visible damage was observed on the surface of the suture piece.However, the ends were noted to be cut, probably caused by a surgical instrument.This product code y497g contains an absorbable suture.Since the sample was received open, the functional test could not be performed due to undetermined exposure time to the environment.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.Additional h3 investigational summary: the product was returned to ethicon for evaluation.One sealed box and one open box with twelve unopened samples each were received for analysis.Product code y497g.As per the sampling plan, a visual inspection was performed on thirteen samples, and no defects were found on the packages.The packets were opened, and the swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along the strand and no issues related to breakage sutures or anomalies were observed during the evaluation.Also, a functional test was performed using instron equipment and tensile force was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained: 1.Any patient consequences? no.2.Were there any unexpected outcomes or complications as a result of the prolonged surgery time? no.3.Was there any change in the patient¿s post-operative care due to the prolonged procedure? no.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|