MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT Back to Search Results
Catalog Number AVSM07080
Medical Device Problem Codes Loss of or Failure to Bond (1068); Positioning Failure (1158); Misfire (2532)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/29/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the delivery system was returned for evaluation in activated condition; the covered stent was found partially deployed upon sample receipt.The slide block, a force transmitting component was no longer connected to the proximal sheath.The disconnection of the slide block is considered to have led to the reported impossibility to completely deploy the covered stent.It was reported that an 8fr introducer sheath was used for access, the tracking vessel was neither tortuous nor calcified and the device was flushed before use.Based on evaluation of the sample, the investigation is closed with confirmed results for bonding joint failure and partial deployment is considered a cascading event.A definite root cause for the reported event could not be established.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.Based on the instructions for use supplied with this product delivery system specific events that could be associated with clinical complications include but are not limited to failure to deploy and high deployment forces.Regarding correct deployment the instructions for use states: "maintain a stationary hold on the white stability sheath during covered stent deployment.Hold the white stability sheath as close as possible to the introducer without touching the dark brown moving catheter of the distal catheter assembly.Maintain the remainder of the white stability sheath relaxed and avoid tension." based on the instructions for use material required for a procedure using the cover a vascular covered stent are 0.035 inch guidewire of appropriate length , introducer sheath with appropriate inner diameter.Regarding preparation the instructions for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated." h10: d4 (expiry date: 09/2025) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a stent graft placement procedure in the left arm fistula graft, the device allegedly had did not deploy.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA VASCULAR COVERED STENT
Common Device Name
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key19792330
Report Number9681442-2024-00195
Device Sequence Number19869886
Product Code PFV
UDI-Device Identifier00801741106255
UDI-Public(01)00801741106255
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P170042
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAVSM07080
Device Lot NumberANHW3242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2024
Initial Report FDA Received Date07/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-