| Model Number |
MI1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that after an accident, the patient was no longer able to hear with the device.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
It was reported that after an accident, the patient was no longer able to hear with the device.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the device investigation revealed mechanical damage to the stimulator electronics that is typical for severe external impact to the housing.The problems described in the recipient report and the reported accident appears to match the damage found.This is a final report.
|
| |
|
Search Alerts/Recalls
|