• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC SMART PORT CT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC SMART PORT CT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number CT75STSD
Medical Device Problem Codes Detachment of Device or Device Component (2907); Infusion or Flow Problem (2964); Separation Problem (4043)
Health Effect - Clinical Codes Numbness (2415); Foreign Body In Patient (2687)
Date of Event 05/22/2024
Type of Reportable Event Malfunction
Event or Problem Description
An end user reported an issue with a single titanium ct vtx port w silc cath.The patient was experiencing left hand numbness and difficulty infusing through the port; however, the decision was made to explant the device.During removal of the port, only about 5 cm of the catheter was able to be removed.A c-arm view showed the line remained, just inferior to the clavicle.The patient was transferred to ir and had the remaining catheter removed on the same day.It is unknown if the device fractured prior to or during the removal procedure.
 
Additional Manufacturer Narrative
The reported device is not available to be returned to the manufacturer for evaluation.An investigation into the root cause for the event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
The customer's reported complaint description of "the patient was experiencing left hand numbness and difficulty infusing through the port; however, the decision was made to explant the device.During removal of the port, only about 5 cm of the catheter was able to be removed.A c-arm view showed the line remained, just inferior to the clavicle.The patient was transferred to ir and had the remaining catheter removed on the same day.It is unknown if the device fractured prior to or during the removal procedure." cannot be confirmed since no complaint sample was returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.Device history review of the packaging, assembly and component lots revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Labeling review: the instructions for use, (16608102-01) which is supplied to the user with this item number, contains the following statements: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.If the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Contraindications catheter insertion in the subclavian vein medial to the border of the first rib, an area which is associated with higher rates for pinch-off.Potential complications: catheter fragmentation, catheter pinch-off, drug extravasation (leakage).Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.The following statements reference making the catheter connection to the port housing: insert catheter into sheath.Position the distal end of the catheter at the desired location.Peel away sheath while withdrawing it from vessel.Care should be taken not to withdraw catheter as sheath is removed.Catheter position should be confirmed radiographically.Secure catheter in place.Trim proximal end of catheter and advance through subcutaneous tunnel to the port pocket.Attach catheter to the port body.Slide the blue strain relief mechanism over the end of the catheter.The tapered end of the blue strain relief mechanism should point away from the proximal end of the catheter.For optimal results, the proximal end of the catheter should be dry.Slide the trimmed end of the catheter tip onto the stem until the catheter is flush with the stem flanges.Slide the strain relief mechanism over the catheter and onto the stem until it contacts the port body.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART PORT CT
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key19793709
Report Number1317056-2024-00113
Device Sequence Number7089007
Product Code LJT
UDI-Device IdentifierH787CT75STSD0
UDI-PublicH787CT75STSD0
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K062414
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Catalogue NumberCT75STSD
Device Lot Number5776465
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2024
Supplement Date Received by Manufacturer09/18/2024
Initial Report FDA Received Date07/22/2024
Supplement Report FDA Received Date09/30/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/25/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
Patient SexFemale
Patient Weight223 KG
-
-