• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 03-2622-3
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 07/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
A user facility's biomedical technician (bmt) reported to customer service that a combi set's heparin line disconnected from the blood pump segment during a patient's hemodialysis (hd) treatment.The patient's estimated blood loss (ebl) is unknown.The complaint device is not available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.Should additional information become available, the file will be updated accordingly.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Additional Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.However, a photograph of the combi set was provided by the customer.The photograph shows blood leaking from the red ¿t¿ connector on the bloodline, where the heparin line connects.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.The reported event was confirmed based on the provided photograph.
 
Event or Problem Description
A user facility¿s biomedical technician (bmt) reported to customer service that a combi set¿s heparin line disconnected from the blood pump segment during a patient¿s hemodialysis (hd) treatment.The patient¿s estimated blood loss (ebl) is unknown.The complaint device is not available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.Should additional information become available, the file will be updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMBI SET
Common Device Name
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key19793731
Report Number0008030665-2024-00638
Device Sequence Number12562578
Product Code FJK
UDI-Device Identifier00840861100231
UDI-Public00840861100231
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K962081
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 08/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number03-2622-3
Device Lot Number23PR01035
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2024
Supplement Date Received by Manufacturer08/06/2024
Initial Report FDA Received Date07/22/2024
Supplement Report FDA Received Date08/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-