| Catalog Number |
03-2622-3 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
07/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Event or Problem Description
|
|
A user facility's biomedical technician (bmt) reported to customer service that a combi set's heparin line disconnected from the blood pump segment during a patient's hemodialysis (hd) treatment.The patient's estimated blood loss (ebl) is unknown.The complaint device is not available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.Should additional information become available, the file will be updated accordingly.
|
| |
|
Event or Problem Description
|
|
A user facility¿s biomedical technician (bmt) reported to customer service that a combi set¿s heparin line disconnected from the blood pump segment during a patient¿s hemodialysis (hd) treatment.The patient¿s estimated blood loss (ebl) is unknown.The complaint device is not available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.Should additional information become available, the file will be updated accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.However, a photograph of the combi set was provided by the customer.The photograph shows blood leaking from the red ¿t¿ connector on the bloodline, where the heparin line connects.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.The reported event was confirmed based on the provided photograph.
|
| |
|
Search Alerts/Recalls
|