| Catalog Number |
760021 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
1 of 1 devices were returned for evaluation.Evaluation of one device found excessive wear due to a lack of proper maintenance.The device did not heat up during evaluation.The device was repaired and returned to the customer.This event is being submitted as part of vmsr.Only one event was received for this device during this quarter.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction events where a rhino xe u-style adapter 7:1 e-type handpiece overheated.No injury resulted.
|
| |
|
Search Alerts/Recalls
|