• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC SIGMARP STB TB IN 3 20.0; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC SIGMARP STB TB IN 3 20.0; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 962135
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 06/29/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient was revised and poly exchange due to infections.Surgical delay was unknown.Doi: (b)(6)2024.Dor: (b)(6) 2024.Affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.A manufacturing record evaluation (nc search) was performed for the finished device product code:962135, lot: 4037908 and no non-conformances / manufacturing irregularities were identified as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot a manufacturing record evaluation (nc search) was performed for the finished device pproduct code:962135, lot: 4037908, and no non-conformances / manufacturing irregularities were identified.H11 additional narrative:.
 
Event or Problem Description
Additional information was received: a.Was there any surgical delay related to the event? no.B.Was there any specific allegation against the reported product? no.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIGMARP STB TB IN 3 20.0
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19794532
Report Number1818910-2024-15553
Device Sequence Number4198071
Product Code NJL
UDI-Device Identifier10603295241287
UDI-Public(01)10603295241287(17)280229(10)4037908
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number962135
Device Lot Number4037908
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2024
Supplement Date Received by Manufacturer07/30/2024
Initial Report FDA Received Date07/22/2024
Supplement Report FDA Received Date07/31/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexFemale
-
-