| Catalog Number |
MC1331P |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a roi-c implant fractured intra-operatively while inserting the anchoring plate.It was reported that the plate was stiff and caused difficulty.The surgery was completed using another cage and plate without patient impact.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d9 and h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the implant has fractured.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to excessive forces applied during anchor insertion.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference report 3004788213-2024-00061.
|
| |
|
Event or Problem Description
|
|
It was reported that a roi-c implant fractured intra-operatively while inserting the anchoring plate.It was reported that the plate was stiff and caused difficulty.The surgery was completed using another cage and plate without patient impact.This is report one of two for this event.
|
| |
|