Catalog Number 244000568
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.Investigation summary: it was reported, that the graters are too dull.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the device found, rounding of cutting edges.It is not unreasonable, that the condition identified in visual analysis would contribute to a dull condition.Therefore, we are able to confirm, dullness with damage observed, as this kind of evidence, indicates repeated use of the device.The lifecycle requirements of the device are event related and depend on the use in clinical practice.The device must be properly inspected, prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions and inspection procedures.A dimensional inspection was not performed, since it was not applicable to the complaint condition.The overall complaint was confirmed.As the observed, condition of the quickset ace grater head, would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life.And it has been determined, that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
It was reported, that the instrument was dull.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET ACE GRATER HEAD 68MM
Common Device Name
HIP INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19798822
Report Number1818910-2024-15585
Device Sequence Number15334822
Product Code HTO
UDI-Device Identifier10603295124153
UDI-Public(01)10603295124153(11)160518(10)SO2024584
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number244000568
Device Lot NumberSO2024584
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2024
Initial Report FDA Received Date07/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1