| Catalog Number |
110031403 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
07/04/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: medical products: item#: 110031427, standard 40mm diameter bearing; lot#: 66119336 g2: foreign: australia h3: customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial shoulder arthroplasty approximately 2 (2) months ago.While the patient was at work approximately eight (8) weeks post op the patient was leaning on their operated arm causing the poly to disassociate from the humeral tray and dislocate.Subsequently, the patient underwent a revision surgery eighteen (18) days ago.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.(b)(4).D10: medical products: item#: 110030778, +6mm lateral offset 40mm diameter glenosphere; lot#: 64630398r.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g2; g3; g6; h1; h2; h3; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.The root cause of the reported issue is attributed to patient noncompliance.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were corrected: d1, d4, d10, h4.The following sections were updated: b4, b5, d9, g3, g6, h2, h6, h11.D10: standard 40mm diameter bearing cat# 110031427 lot# 66138564.Cr 40mm glenosphere +3mm cocr cat# 110030777 lot# 65942426.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial shoulder arthroplasty approximately one (1) year and four (4) months ago.Subsequently, the patient underwent a revision surgery approximately two (2) months later due to the poly disassociating from the humeral tray and dislocation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g2; g3; g6; h1; h2; h3; h6.Visual examination of the returned tray identified signs of use with wear and tear in the center portion of the tray, around the outside edge of the tray, and on the post and under side of the tray.Product identifiers were found conforming to the print specification.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.Further information provided by the surgeon indicated the patient was not leaning on the operated arm.This information was provided in error.As such, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|