MAUDE Adverse Event Report: ZIMMER BIOMET, INC. MINI TRAY +5MM COCR +3 OFFSET; SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. MINI TRAY +5MM COCR +3 OFFSET; SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES Back to Search Results
Catalog Number 110031403
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 07/04/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: medical products: item#: 110031427, standard 40mm diameter bearing; lot#: 66119336 g2: foreign: australia h3: customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent an initial shoulder arthroplasty approximately 2 (2) months ago.While the patient was at work approximately eight (8) weeks post op the patient was leaning on their operated arm causing the poly to disassociate from the humeral tray and dislocate.Subsequently, the patient underwent a revision surgery eighteen (18) days ago.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.(b)(4).D10: medical products: item#: 110030778, +6mm lateral offset 40mm diameter glenosphere; lot#: 64630398r.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g2; g3; g6; h1; h2; h3; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.The root cause of the reported issue is attributed to patient noncompliance.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were corrected: d1, d4, d10, h4.The following sections were updated: b4, b5, d9, g3, g6, h2, h6, h11.D10: standard 40mm diameter bearing cat# 110031427 lot# 66138564.Cr 40mm glenosphere +3mm cocr cat# 110030777 lot# 65942426.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Event or Problem Description
It was reported that the patient underwent an initial shoulder arthroplasty approximately one (1) year and four (4) months ago.Subsequently, the patient underwent a revision surgery approximately two (2) months later due to the poly disassociating from the humeral tray and dislocation.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g2; g3; g6; h1; h2; h3; h6.Visual examination of the returned tray identified signs of use with wear and tear in the center portion of the tray, around the outside edge of the tray, and on the post and under side of the tray.Product identifiers were found conforming to the print specification.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.Further information provided by the surgeon indicated the patient was not leaning on the operated arm.This information was provided in error.As such, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI TRAY +5MM COCR +3 OFFSET
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19800415
Report Number0001822565-2024-02439
Device Sequence Number18218298
Product Code PHX
UDI-Device Identifier00887868231353
UDI-Public(01)00887868231353(17)291031(10)64550813
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K181611
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number110031403
Device Lot Number64550813
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2024
Supplement Date Received by Manufacturer10/02/2024
10/30/2024
02/20/2025
09/24/2025
Initial Report FDA Received Date07/22/2024
Supplement Report FDA Received Date10/24/2024
11/08/2024
03/18/2025
09/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/07/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11; SEE H11 NARRATIVE; SEE H11 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexMale
Patient Weight120 KG
-
-