• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number RLT351416
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Aneurysm expansion (4597)
Date of Event 12/01/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Additionally, it was reported by the physician that the patient has discontinued follow-up with vascular surgery secondary to her age being (b)(6) years old.The type endoleak is reported to be a type i-a and type ii endoleak.
 
Additional Manufacturer Narrative
B7: patient medical history includes but is not limited to: carotid disease, coronary artery disease, hypercholesterolemia, hyperlipidemia, congestive heart failure, coronary artery bypass graft, chronic obstructive pulmonary disease, diabetes mellitus, h6: code 22: the instructions for use (ifu) for the gore® excluder® aaa endoprosthesis states; adverse events that may occur and / or require intervention include but are not limited to: endoleak, aneurysm enlargement w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
On march 24, 2015, this patient underwent endovascular treatment of an abdominal aortic aneurysm with a maximum aortic diameter measuring 50mm; and was implanted with gore® excluder® aaa endoprostheses featuring c3 delivery system.The patient tolerated the procedure and was discharged from the hospital on (b)(6) 2015.On (b)(6) 2015, follow-up computed tomography angiography (cta) determined the maximum diameter of the aneurysm measured 47.5mm.On (b)(6), 2020, follow-up cta determined the maximum diameter of the aneurysm measured 58.5mm.On (b)(6), 2021, it was determined that the patient had a 12.3mm increase in endoleak size since 2015.There was no additional procedure recorded for the patient and the endoleak type was not designated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY MOUNTAIN B/P
3750 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
laura crawford
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key19800703
Report Number2017233-2024-05130
Device Sequence Number15302360
Product Code MIH
UDI-Device Identifier00733132618484
UDI-Public00733132618484
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2017
Device Catalogue NumberRLT351416
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/22/2024
Supplement Report FDA Received Date07/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/20/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient Age89 YR
Patient SexFemale
Patient Weight89 KG
-
-