• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL DELTAFILL18 6MM X 25CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL DELTAFILL18 6MM X 25CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number DLF180625
Medical Device Problem Code Stretched (1601)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/05/2024
Type of Reportable Event Malfunction
Event or Problem Description
The healthcare professional reported that during a coil embolization of the axillary artery via simple technique, the complaint coil, a 6mm x 25cm deltafill 18 (dlf180625 / 31178127) was used for filling.It was reported that the coil became unraveled while being inserted and removed several times.As a result, the coil was removed along with the concomitant microcatheter, an excelsior® 1018® microcatheter (stryker).After the microcatheter was inserted into the target lesion again, additional coils were implanted and the procedure was completed without any negative patient impact.Continuous flush was maintained through the microcatheter.On 09-jul-2024, additional information was received which indicated that the coil had been withdrawn and repositioned several times.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, sex, gender, weight, race, and ethnicity were not provided.Section d.2b: procode is krd/hcg.Section e.1: the initial reporter phone: (b)(6).The initial reporter email address is not available / reported.Based on complaint information, the device is not available to be returned for analysis.A review of manufacturing documentation associated with this lot (31178127) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the information available and without the complaint product available to be returned for analysis, the reported product issue documented in the complaint cannot be confirmed through functional evaluation and analysis.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint device; however, it is possible that clinical and procedural factors, including device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTAFILL18 6MM X 25CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC. (FREMONT)
47709 fremont blvd
fremont CA 94538
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
303552-6892
MDR Report Key19801504
Report Number3008114965-2024-00672
Device Sequence Number7496855
Product Code KRD
UDI-Device Identifier10886704076933
UDI-Public(01)10886704076933(11)231110(17)261031(10)31178127
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDLF180625
Device Lot Number31178127
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2024
Initial Report FDA Received Date07/22/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR® 1018® MICROCATHETER (STRYKER).
Patient SexUnknown
-
-