EXACTECH, INC. OPTETRAK LOGIC; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-45-3040 |
| Medical Device Problem Codes |
Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Granuloma (1876); Joint Laxity (4526); Joint effusion (4849)
|
| Date of Event |
02/27/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a male patient, initial left knee implanted on (b)(6) 2019, underwent a revision procedure on (b)(6) 2024, approximately 4 years 8 months post the initial procedure.The patient was suffering from laxity, instability, and chronic hydrarthrosis.Macroscopic pe granuloma indicated.The report stated the patient had aseptic tibial and femoral loosening 5 years after the initial placement.The pe was changed.No surgical delays or device breakages were reported.No x-rays were provided.The device return status for analysis is unknown currently.They have been requested.No device images were provided.No further information.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: concomitants: 5893383 - 200-02-41 - implant patellaire 3 plots, 5705289 - 204-34-04 - tige extension diam 14 lg 40, 5743666 - 04-70-00 - vis pour tige extension tibiale, 2028019157 - a10012 - gps implant kit v2,.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported was likely the result of prosthesis wear and insufficient bonds between the components and the bones, which led to aseptic (non-infected) femoral and tibial loosening.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.
|
| |
|