• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. IPC® HANDPIECE - XPS® STRAIGHTSHOT® M4; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. IPC® HANDPIECE - XPS® STRAIGHTSHOT® M4; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I Back to Search Results
Model Number 1898200T
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that pre-operative, m4 handpiece had heating issue.There was no patient involved.
 
Additional Manufacturer Narrative
H3:product analysis could not perform temperature test as ipc showing error code 15.Connector has dent and cable damaged.Thumb wheel jammed.Hi-pot test fail.Blade/bur not rotating.Motor cable assembly found faulty.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803." medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information received stated that the heating was observed after 2 minutes of use.The handpiece was being run a 3000 rpm in oscillate mode with no irrigation being used.The procedure was completed manually.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPC® HANDPIECE - XPS® STRAIGHTSHOT® M4
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key19803962
Report Number1045254-2024-01107
Device Sequence Number8303594
Product Code ERL
UDI-Device Identifier00643169350687
UDI-Public00643169350687
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K041413
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1898200T
Device Catalogue Number1898200T
Device Lot Number208404771
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2024
Supplement Date Received by Manufacturer08/04/2024
Initial Report FDA Received Date07/23/2024
Supplement Report FDA Received Date08/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-