| Model Number |
GIF-XP190N |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device inspection, the gastrointestinal videoscope exhibited foreign material clogged in the nozzle.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 years since the subject device was manufactured.Based on the results of the investigation and the information provided, it could not be determined what the foreign material was.It was likely caused by insufficient reprocessing due to leakage from the connector contact pin.However, a definitive root cause could not be determined.The instruction manual states the detection method associated with the event in "gif/cf/pcf-190 series operation manual chapter 3 preparation and inspection", and preventative measures in "gif/cf/pcf-190 series reprocessing manual chapter 5 reprocessing the endoscope".Three attempts were performed to obtain additional information, but no response was received from the customer.Olympus will continue to monitor field performance for this device.
|
| |
|