The manufacturer received information related to a simplygo mini oxygen concentrator.The device was returned to a third-party service center.There is no allegation of patient harm, serious injury, or medical intervention.The device was returned to an authorized service center for evaluation.The device was visually inspected.During the evaluation the service center found pca was burned.In addition, sieve canister is clogged, o2 side is defective, air side and inlet filter are contaminated.Lastly all test result passed.
|