SYNTHES GMBH LOCKING SCREW FOR IM NAIL 5/ 66/ XL25; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 04.045.066
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 06/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2024 patient was brought into surgery by surgeon for a loosening of distal interlocking screws in a synthes rfna femoral nail.The implant was initially placed by surgeon on (b)(6) 2024 at the same facility.At some point after that date the patients nursing home noticed prominent hardware at the skin.The patient was brought to the hospital for revision surgery on (b)(6) 2024.There was an infection.There was no surgical delay.This report involves one (1) locking screw for im nail ø 5mm/ 66mm/ xl25.This is report 2 of 6 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D9: complainant part is not expected to be returned for manufacturer review/investigation.D10 therapy date: (b)(6) 2024 h3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: a1: patient identifier added.H3, h6: the product was not returned to depuy synthes, however x-ray were provided for review.The x-ray investigation revealed nothing indicative of a device nonconformance.No problem was detected.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the locking screw for im nail 5/ 66/ xl25 would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.The device lot number is unknown, therefore a device history review could not be performed.If more information become available, the record will be re-assessed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW FOR IM NAIL 5/ 66/ XL25
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19806114
Report Number8030965-2024-09066
Device Sequence Number7387135
Product Code HWC
UDI-Device Identifier10886982292964
UDI-Public(01)10886982292964
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K201346
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.045.066
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/27/2024
Supplement Date Received by Manufacturer08/14/2024
Initial Report FDA Received Date07/23/2024
Supplement Report FDA Received Date08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LOCKING SCREW FOR IM NAIL 5/ 38/ XL25; LOCKING SCREW FOR IM NAIL 5/ 82/ XL25; LOCKING SCREW FOR IM NAIL 5/ 90/ XL25; RFNA/ 12MM/ 360MM/ STANDARD BEND/ STER; RFNA/ 12MM/ 360MM/ STANDARD BEND/ STER
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-