• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE "TRUEVIEW II", 4 MM, 70°, AUTOCLAVABLE; RIGID ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE "TRUEVIEW II", 4 MM, 70°, AUTOCLAVABLE; RIGID ARTHROSCOPE Back to Search Results
Model Number A70942A
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, the telescope had a bent scope.The issue occurred during a therapeutic procedure.There were no reports of patient harm.Although there was a slight delay, the intended procedure was able to be successfully completed using a similar device.
 
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the results of evaluation and legal manufacturer's final investigation.From the results of evaluation, the following items were observed: - confirmed - bent outer tube.- scratches on distal edge.- optical fiber - no image.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to excessive force such as a fall, impact or similar stress.There are no countermeasures necessary for the suggested phenomenon because the associated risk is acceptable.The manufacturer will monitor the occurrence rate in the context of the utilized quality management system.If necessary, the manufacturer will take action in the future.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE "TRUEVIEW II", 4 MM, 70°, AUTOCLAVABLE
Common Device Name
RIGID ARTHROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19807262
Report Number9610773-2024-01783
Device Sequence Number20919782
Product Code HRX
UDI-Device Identifier04042761010559
UDI-Public04042761010559
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K951354
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA70942A
Device Lot Number808376
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2024
Supplement Date Received by Manufacturer10/24/2024
Initial Report FDA Received Date07/23/2024
Supplement Report FDA Received Date10/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/16/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-