WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G34505
Medical Device Problem Codes Difficult to Remove (1528); Structural Problem (2506); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Dyspnea (1816); Hematoma (1884); Hemorrhage/Blood Loss/Bleeding (1888); Pain (1994); Shock (2072); Perforation of Vessels (2135); Respiratory Failure (2484); Retroperitoneal Hemorrhage (4483); Insufficient Information (4580)
Type of Reportable Event Serious Injury
Event or Problem Description
It is alleged that the patient received a celect platinum (pt) inferior vena cava (ivc) filter on (b)(6) 2023, and the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
 
Additional Manufacturer Narrative
Section e3: non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged celect pt is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: a2, b1, b2, b5, b6, d4, g4, h4, h6.G4 (510k): k211875.Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: vena cava (vc) perforation, open filter retrieval, hematoma, hemorrhagic shock, respiratory failure, repair of artery/vein/inferior vena cava (ivc), shortness of breath (sob), back pain.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: vascular trauma, vena cava perforation, vena cava penetration.Physician practice guidelines and published guidance from regulatory agencies recommend that patients with indwelling filters undergo routine follow-up.The risks/benefits of filter retrieval should be considered for each patient during follow-up.Once protection from pe is no longer necessary, filter retrieval should be considered.Filter retrieval should be attempted when feasible and clinically indicated.Filter retrieval is a patient-specific, clinically complex decision; the decision to remove a filter should be based on each patient¿s individual risk/benefit profile (e.G., a patient¿s continued need for protection from pe compared to their experience with and (or) ongoing risk of experiencing filter-related complications).The filter is designed to be retrieved with the günther tulip vena cava filter retrieval set.It may also be retrieved with the cloversnare® vascular retriever.Cook has not performed testing to evaluate the safety or effectiveness of filter retrieval using other retrieval systems or techniques.The published clinical literature includes descriptions of alternative techniques for filter retrieval; use of these techniques varies according to physician experience, patient anatomy, and filter position.The safety or effectiveness of these alternative retrieval techniques has not been established.Specific for ¿embedded¿ a filter that is embedded in the wall of the ivc may be difficult to retrieve.For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.Unknown if the reported hematoma, hemorrhagic shock, respiratory failure, repair of artery/vein/ inferior vena cava (ivc), shortness of breath (sob), and back pain are directly related to the filter and unable to identify a corresponding failure mode at this point in time.20 devices in lot.No other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Patient allegedly received a cook celect platinum filter via the femoral vein due to pulmonary embolism (pe) and deep vein thrombosis (dvt).Patient is alleging laparoscopic filter retrieval, hemorrhagic shock due to perforation of ivc and lumbar artery.It is also alleged the patient presented to the emergency room (er) with complications of back pain and shortness of breath.Patient was found to have a large retroperitoneal hematoma secondary to ivc filter perforating the ivc status post exploratory laparotomy, ligation of lumbar artery, repair of the ivc and right renal vein, venotomy and ivc filter extraction.Patient was treated for hemorrhagic shock and acute respiratory failure.Filter retrieval date is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET
Common Device Name
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
henriette s. christiansen
sandet 6, dk-4632
bjaeverskov 
MDR Report Key19807865
Report Number3002808486-2024-00135
Device Sequence Number20707115
Product Code DTK
UDI-Device Identifier10827002345055
UDI-Public(01)10827002345055(17)260621(10)E4420830
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2026
Device Model NumberG34505
Device Catalogue NumberIGTCFS-65-1-UNI-CELECT-PT
Device Lot NumberE4420830
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/26/2024
Supplement Date Received by Manufacturer10/09/2024
Initial Report FDA Received Date07/23/2024
Supplement Report FDA Received Date11/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention; Hospitalization;
Patient SexMale
-
-